Cat. COR-70VET-WM Corix Pro 70 Dual Mode Wall-Mounted Veterinary X-Ray Equipment
Description:
Wall mountable, equipped with a long-range driftless positioning arm. It could be installed for use in walk-through applications, allowing for more operating positions. By replacing the short intra-oral diagnostic cone with our exclusive Beam Centering Device Mod. Q100 (Optional) Will produce quality radiographic images collimated to the physical dimensions of a standard radiographic cassette, or flat panel CCD sensor, enabling full-body veterinary diagnosis of small and medium-weight species.
Technical Specifications:
- Supply voltage: 120-127-230 V. +/- 10%.
- Power consumption: 800 VA @ 120 V.
- Tube potential: 70 KVp +/- 7% (single-phase, self-rectifying).
- Anode Current: 8 mA +/- 15% @ nominal line voltage.
- Exposure time range: 0.03 to 3.00 s., In steps of 0.01 s. 10 preset exposure times, with automatic main voltage compensation. 5 different pre-selected pre-heating times by the user are available for loading the X-ray tube. (Pre-heating time is not shown on the screen).
- Minimum focal length for dental x-rays: 200 MM. (8").
- Focal spot: 0,8 (lEC 60366: 2005).
- Total filtration: 2 mm. To the eq.
- Leak radiation: Less than 50 mR/h. (Technical factors: 70KVp, 8mA, 1s).
- Net weight: Mod. WM (wall mounting): 29 kg.
- Shipping Weight: Mod. WM (wall mount): 34 kg. .
- Shipping dimensions: Mod. WM (Wall Mount): 1 Box, 95 x 52 x 28 cm.
Ratings